MiniSim 1000 Patient Simulator
ECG Normal Sinus:
Rate: |
30, 60, 70, 80, 90, 100,120, 150, 180, 210, 240, 270, 300, 350 BPM |
Amplitude: |
0.15 to 5mV, High Level |
Output: |
0.5 Volt/mV of Low Level (Lead II) |
Test Waveforms:
Sine, Triangle, Square Rate: 0.1 to 100 Hz |
0.1 to 1 Hz in 0.1 Hz increments |
1 to 10 Hz in 1 Hz increments, 10 to 100 Hz in 10 Hz increments |
Tall T Wave:
Test Waveform@ 70 BPM, Atrial Pacer, Ventricular Pacer*, Paced Rhythms - Variable from 0.1to 2mS, R Wave detection - Variable form 10 to 120 ms |
Pulse: 2mS Pulse of 1mV amp @ 4 sec intervals |
Arrhythmias:
ST Segment Analysis, ST Elevation and Depression, ST Depression Positive and Negative Slope, Myocardial Infarction, Atrial, Ventricular, Blocks |
Respiration:
Normal Physiological Simulation: |
Base Line |
Impedance: |
250, 500, 750, 100 Ohms |
Delta Impedance: |
0.1, 0.5, 1.0, 1.5 Ohms |
Rates: |
15, 30, 60, 120 BPM |
Apnea: |
Off, Continuous, 12 Seconds and 32 Seconds Apnea-Bradycardia |
Accuracy: |
Delta Impedance 5% of selection, Rate: 1% of selection |
Blood Pressure:
Impedance: |
350 Ohms |
Excitation: |
2 to 16 Volts |
Sensitivity: |
5uV/V/mmHg |
Static: |
0, 5, 10, 20, 25, 30, 40, 50, 100, 150, 200, 300 mmHg |
Dynamic: |
50/10, 60/20, 70/30, 80/40, 100/60, 120/80 mmHg |
Accuracy: |
±2% FS. ±1 mmHg |
Synchronized with all normal sinus rates |
|
Physiologically track all arrhythmia selections |
Temperature:
Compatibility: |
YSI 400/700, 25/37 Degree Centigrade |
Accuracy: |
±1% of reading |
General:
Size: |
5½ x 3½ x 1 ½ inches (13.97 x 8.9 x 3.8 cm) |
Weight: |
10 Oz (0.28 kg) |
Case: |
High Impact Plastic |
Environmental:
Operating Temperature: |
59° to 95° F (15° to 35° C) |
Storage Temperature: |
32° to 131° F (0° to 55° C) |
Netech Minisim 1000 is a compact, multiparameter physiological waveform simulator. It simulates waveforms such as ECG, IBP, respiration, temperature. It is designed to test and verify the proper functionality of the patient monitors.
This device is an ideal all-in-one solution that can simulate ECG, arrhythmia, blood pressure, respiration, and temperature. It goes the extra mile for you by also being able to replicates performance waveforms including sine, triangle, pulse and square wave.
Full 12 lead ECG simulation with 14 user selectable rates from 30 to 350 BPM and 14 user selectable amplitudes from 0.15 to 5 mV are all available. Sine, square, and triangle performance waveforms are simulated with 14 user selectable rates from 0.1 to 100 Hz.
In addition to ECG simulation, it simulates 45 different arrhythmias including atrial, ventricular, tall T wave, ST elevation, ST depression, myocardial infarction, blocks, pulse, and pacer waveforms. In addition, two invasive blood pressures are simulated with a choice of 12 static pressures and 6 dynamic waveforms.
Respiration is simulated with 4 rates, 4 baseline impedance, and 4 delta impedance selections. Apnea may be off, continuous, and with either 12 or 32-second apnea bradycardia. Finally, the temperature is available at either 25° and 37° C (77° and 98.6° F). (Optional pressure and temperature interface cables are available for most manufacturers.)
Powered by a 9-volt alkaline battery with a typical life of 40 continuous hours, the MiniSim 1000 provides an unmatched variety of patient simulations in an efficient, well-designed package. It is an incredibly powerful tool that’s built for on-the-go testing, and it eliminates the need for you to carry and learn to operate multiple devices.
FEATURES:
- 12 lead ECG
- 45 arrhythmias
- ECG performance waveforms
- 2 blood pressure simulations
- Respiration / Apnea
- Temperature
- Accurate, portable, and easy-to-use
- Menu-driven operation
- Large LCD Display
- Tactile Keypad
- Battery or AC Powered
- Product Code: 300-1.3
- Availability: In Stock